New TopShim run:

version = topshim 2.0.0
date = 2023-10-27
time = 23:32:21.957
command = topshim  convcomp tuneaxyz

1D SHIMMING

Parameters:

maximum order = 5
probe = Z116098_0395
solvent = CDCl3
shim nucleus = 2H
nucleus optimised for = 13C
o1p (from lock) = 7.24 ppm
optimisation parameters = ls
linewidth 1H = 0.10 Hz
envelope shape / strictness = 2.00 / 200

Results:

tune lock level increase = 1.3

initial B0 stdDev = 5.92 Hz
sample size = 2.70 cm, position = 0.10 cm
final B0 stdDev = 1.09 Hz > improvement = 5.4
envelope width = 0.89 Hz

Performing tune after shimming
tune lock level increase = 1.1

shim changes: 
Z	+103
Z2	-59
Z3	+158
Z4	-602
Z5	+180
X	-55
Y	+130

duration = 4 min 2 sec
completed successfully
finished Fri Oct 27 23:36:22 2023


========================================================================